மருந்து அறிவியல் மற்றும் வளர்ந்து வரும் மருந்துகளின் இதழ்

Formulation Characterization of Parathyroid Hormone PTH (1-34)Coated on a Novel Transdermal Microprojection Delivery System:Water Determination

Mahmoud Ameri, Shelley C Fan and Yuh Fun Maa

This study assessed the feasibility of determining the water content of the PTH (1-34) formulation coated on a novel transdermal microprojection delivery system, ZP-PTH. Thermogravimetric analysis (TGA) and Karl Fischer (KF) titration were used for water analysis.Headspace moisture was determined by internal vapor analysis and a hand-held moisture sensor. The moisture sorption isotherm was established by moisture sorption analyzer. PTH (1-34) API and formulation showed several weight-loss phases in TGA thermograms. Water loss is not a single-phase event, occurring over a wide temperature range. The typical water amount in a single ZP-PTH coated array is below the sensitivity of TGA and KF titration. Combining 20 coated array samples together in a single extract made KF analysis is possible but impractical due to strenuous sample preparation requirements and controls. The moisture sorption isotherm allowed accurate prediction of water content when the headspace moisture level is known. We demonstrated that TGA and KF titration are not practically possible for determining the water content in ZP-PTH. Simple alternative approaches can ensure that the desired water content in the ZP-PTH coating is achieved without direct water measurement. This study provides scientific rationales to justify the application of unconventional approaches.

மறுப்பு: இந்த சுருக்கமானது செயற்கை நுண்ணறிவு கருவிகளைப் பயன்படுத்தி மொழிபெயர்க்கப்பட்டது மற்றும் இன்னும் மதிப்பாய்வு செய்யப்படவில்லை அல்லது சரிபார்க்கப்படவில்லை